Specificity

Fit your standard curve, then compare a clean reference sample against samples spiked with potential interfering substances, to get FDA Bioanalytical Guidance (2018) Table 1 specificity verdicts for ligand-binding assays — each interferent within ±20% of the reference (±25% if tested at the LLOQ/ULOQ level).

1. Standard curve

Paste calibrator concentration + OD pairs, one per line (tab/comma/space separated). Extra columns are treated as OD replicates and averaged for the fit, e.g. 1.25, 0.44, 0.43, 0.46. The concentration column doubles as each calibrator’s nominal value for the back-calculation check below.

2. Specificity data

Enter the OD of a clean reference sample (no interferent), then the OD of each sample spiked with a potential interfering substance at a relevant concentration. Each interferent is compared against the reference sample’s own back-calculated concentration — not a separately-entered nominal.

InterferentOD