Selectivity
Fit your standard curve, then paste OD readings from at least 10 individual matrix sources spiked at LLOQ and at a high QC concentration, and get FDA Bioanalytical Guidance (2018) Table 1 selectivity verdicts for ligand-binding assays — each source within ±25% of nominal at LLOQ and ±20% of nominal at the high QC level.
1. Standard curve
Paste calibrator concentration + OD pairs, one per line (tab/comma/space separated). Extra columns are treated as OD replicates and averaged for the fit, e.g. 1.25, 0.44, 0.43, 0.46. The concentration column doubles as each calibrator’s nominal value for the back-calculation check below.
2. Selectivity data
Enter the LLOQ and high-QC nominal concentrations, then paste each individual matrix source’s OD when spiked at LLOQ and at the high QC level. FDA Table 1 wants ≥10 individual sources — fewer is allowed here for testing, but the report will flag it.
| Source | OD at LLOQ spike | OD at High QC spike |
|---|